Obtura Spartan · BUC-E® ultrasonic tips
The right tip on the wrong thread damages the scaler. The BUC® 1E is the BUC® 1 access-refinement tip cut with the M3 x 0.5 thread – the E series, which Obtura Spartan describe as the EMS fitting. Same 0.6mm diamond-coated rounded end, which Obtura state eliminates clefts and ditches in the pulp chamber floor.
This pack: Obtura Spartan BUC® 1E "The Cutter" Ultrasonic Tip, 0.6mm diameter · single tip · diamond coated · supplied non-sterile
› Working end: 0.6mm, fine grit, diamond coated. Water port at the working end.
› Thread: M3 x 0.5 – the E series, Obtura Spartan's EMS fitting
› Intensity: Medium to high – Obtura Spartan's guidance
› Cutting and refinement of line angles
› Chasing calcifications
› Smoothing access walls
Supplied non-sterile. Reprocessing is validated to ISO 17664-1 and ISO 17665-1: pre-clean in a pH-neutral ultrasonic solution, dry, then prevacuum steam at 132°C for 4 minutes.
Will it fit my ultrasonic unit? Obtura Spartan make the range for scalers using M3 x 0.6 threading, and offer every tip again as the E series with M3 x 0.5. This one is M3 x 0.5. Check your unit's manual first – they warn that forcing a tip with the wrong thread damages the device. The tip wrench seats it properly.
How long does a tip last? Obtura Spartan state the tips were tested to 250 reprocessing cycles to ISO 18397 without loss of performance or corrosion – but end of life is set by wear in use as well, so inspect each tip during cleaning and retire damaged ones. They also ask for very light pressure: small-diameter tips are subject to breakage at any time.
| Fitting and storage | Wrench for Spartan Tips · Spartan Tip Stand |
| Patency and irrigation | C-PILOT® 008 25mm · Chlorax 5.25% · Endo-Solution 17% EDTA |
| Rest of the range | BUC® 2E Planer · BUC® 3E Digger · BUC® 1, M3 x 0.6 |
Diamond coated · Fine grit · M3 x 0.5 · Autoclavable 132°C · REF 932-018
SpotlightRECIPROC® MinimaOne file · slim shaping for narrow canals
Shaping systems
Hand files & finders