Cerkamed Endo-Top · 30G side-vented · 25 mm · 100
Canal irrigationPeriodontal pocketsClosed end, side ventSterile, single use
Endo-Top is Cerkamed's irrigation needle: 30G, 25 mm, closed rounded tip with a side vent, so the irrigant leaves sideways at low pressure instead of straight down the canal. Cerkamed's wording: "The product acts by on irrigation using a turbulent flow of fluid, which is dispensed through the side hole of the needle causing low pressure on the tissues." Luer lock, ethylene-oxide sterilised, one hundred to the box.
This pack: Cerkamed Endo-Top Irrigation Needles, 30G (0.3 mm) x 25 mm, closed end, side-vented, Luer lock, sterile (Pack of 100)
| Gauge and length | 30G – Cerkamed's metric designation 0.3 mm x 25 mm |
| Tip | Closed, rounded end with a lateral vent |
| Fitting | Luer-lock thread, for a standard irrigation syringe |
| Sterility | Sterilised with ethylene oxide, single use, individually blistered |
| Use | Cerkamed: rinsing root canals and gingival pockets, with root canal rinsing solutions only |
| Pack | 100 |
› Cerkamed's three steps: "Open the blister. With a Luer-lock thread, mount the needle onto a syringe filled with the chosen fluid. Remove the needle sheath and check the needle for visible defects or fractures." Then expel a little fluid to confirm the flow before it goes in the canal.
› From the IFU, the depth the web page leaves out: "measure its length and then set the stop (endostop) on the needle to a depth of 1-1.5 mm less than the measured working length. Inserting the needle to this depth minimises the risk of it getting stuck in the canal."
› Cerkamed's IFU on the delivery: "The liquid should be introduced into the root canal with continuous pumping movements." And: "do not use too strong a liquid jet, as this may cause the liquid to penetrate the periapical tissues. If resistance is encountered during needle insertion, the needle should be carefully withdrawn."
› For a pocket rather than a canal, Cerkamed's IFU: "the liquid jet should be directed at the surface of the tooth and not at the bottom of the pocket."
› Cerkamed's IFU: "Avoid excessive twisting and bending, which may disturb the integrity and cause the needle to break. Use with luer-lock syringes only."
› Cerkamed: "Perform root canal rinsing according to instructions given by manufacturer of chosen rinsing solution. Needles are intended to be used only with the liquids for root canal rinsing."
› Cerkamed: "Re-use poses a risk of infection. Do not reuse." And if the blister is damaged or opened, "the product should not be used because sterility is no longer guaranteed".
› Cerkamed note the product "may contain residues from ethylene oxide sterilisation at an acceptable level according to EN ISO 10993-7".
Because an open-ended needle pushes the irrigant apically and a side vent pushes it at the wall. Cerkamed describe the effect as turbulent flow at low pressure on the tissues. That is the design argument; how far short of the apex to set the stop is the clinician's call and the irrigant manufacturer's directions.
A 0.3 mm cannula is what Cerkamed make it, and 25 mm is the working length of most syringes it will sit on. For a canal that has only been scouted with a C-PILOT ISO 010, 25 mm it may not reach far; once shaped, it does – to 1–1.5 mm short of working length, which is where Cerkamed's IFU has the stop set. Do not pre-bend it: the IFU says to avoid twisting and bending.
| Irrigants | Chloraxid 5.25% · Chloraxid 2% · Endo-Solution 17% EDTA · Citric Acid 40% |
| Chemo-mechanical prep | Endo-Prep Cream 10 ml |
| Activation | EDDY |
| Evacuation | Endo-Aspirator PRO · Endo-Aspirator (5) |
| Other needles | IrriFlex |
Endo-Top · 30G 0.3 x 25 mm · closed end, side-vented · Luer lock · sterile, single use · Pack of 100 · Made by Cerkamed, Stalowa Wola, Poland · REF CER-W-ET100N