Obtura Spartan · Tools
Nothing on the Obtura III Max is meant to be done with fingers or pliers. The Multi-Tool unlatches the thermal protector, turns the needle nut on and off the heater thread, bends the applicator needle, and carries the small hex opening that unscrews the plunger seal assembly. It is the tool every maintenance instruction in the manual assumes you are holding.
This pack: Obtura III Max Multi-Tool · one tool · autoclavable · for the Obtura III Max
› Thermal protector: unlatched and removed with the Multi-Tool at the start of every clean
› Needle nut: unscrewed and re-tightened with it, and the nut stays on the tool while the needle is changed. Obtura Spartan say to use it to assure proper fit
› Bending the needle: “always use the Multi-Tool to bend the needle” – once only, at the largest diameter, as close to the thermal protector as possible
› Plunger seal: the small hex opening is what the seal assembly unscrews into
Because the Multi-Tool contacts potentially contaminated surfaces, Obtura Spartan require it to be cleaned and sterilised before first use and after every subsequent use – the same requirement as the thermal protectors, and an easy one to miss on something that looks like a spanner.
› Clean first: wipe off visible soil, wet with tap water, brush with a soft bristle brush until clean, rinse under potable running water for 30 seconds, dry and inspect
› Then sterilise: in a paper/plastic pouch, prevacuum steam, full cycle 132°C for 4 minutes, in an ANSI AAMI ST55 or ST8 compliant autoclave
› Drying: 30 minutes on the autoclave drying cycle, never above 137°C, and back to room temperature before use
One came with my unit – why buy a second? Turnaround. The tool is sterilised between patients, so a single one means the instrument waits on the cycle. Practices running two surgeries off one Obtura tend to buy a second as a matter of course.
Does it fit the older Obtura II? Obtura Spartan list 823-813 as the Obtura III Max Multi-Tool. If you are running an Obtura II, ask us before ordering.
| The unit | Obtura III Max System |
| What it fits | Needle Nuts (2) · Plunger Seal Assembly · Thermal Protectors (4) |
| Needles it bends | 20G Black (5) · 23G Blue (5) · 25G Green (5) |
| Keeping it clean | Obtura Cleaning Kit · Cleaning Brushes (2) |
Single tool · Autoclavable 132°C / 4 min · Obtura III Max · REF OBT-823813 · mfr code 823-813